COUNTY OF LOS ANGELES SPECIFICATIONS For

Date: 10/14/13 Solicitation #: RFB-IS-1420065

Requisition #: RQN-SH-14017157

For: Fourrier Trans Infra Red Detector (FTIR) BIDDER TO COMPLETE THE FOLLOWING INFORMATION

Company Name: Address: Contact Person: Tel: No.:

Fax No.:

E-mail address: Notice: Bidder shall state in the right hand column wherein your product offered differs, indicating performance, specific size, and/or make and model of all components when not exactly as specified. When bidder is bidding items exactly as described in the left hand column, please state “AS SPECIFIED” on the right hand column. Failure to return and fill in this form will be considered sufficient reason for rejection of your offer. Literature alone is not sufficient for consideration of your offer. All equipment must meet California and County of Los Angeles safety requirements. The equipment shall be the latest model and shall not have been used as a demonstrator. Bidders shall submit detailed literature on the unit they propose to furnish.

Bid #

Page 1 of 4

(Rev. 03/08)

REQUIREMENTS

INDICATE EXCEPTION OR STATE “AS SPECIFIED” BELOW

1.Specifications for one Bruker Alpha FT-IR baser spectrometer or equal. Fourrier Trans Infra Red Base spectrometer must have a spectral resolution better than 2cm-1. Spectral range must be 375-7500 cm-1 wavenumbers. Must be able to provide accurate identification of chemical weapons, explosives, toxic industrial chemicals and narcotics with included libraries. Must ba able to quickly attach sampling modes (must have Quicksnap sampling modes or equal) -RockSolid interferometer (or Equal) with gold mirrors, high throughput and permanenet alignment. Automated instrument tests with integrated polystyrene standard -Detector:DTGS -Source: SiC globar, exchange by user -sealed and desiccated optic, desiccant reclaimable. -PerformanceGuard: continuos monitoring of all spectrometer componenets, performance and humidity -Opus/Mentor Spectroscopy software -Desiccant -Set of tools System shall no larger than 22 cm x 30 cm including sampling module. 2. Quick Snap sampling module A220/D-01 -Platinum-ATR-sampling module, diamond, 1 bounce, Quicksnap to include: -Daimond Crystal, single reflection, edged in tungstencarbide -Work disk, stainless steel -spectral range:375-7,500 cm-1 -Ergtonomic one finger clamp mechanism pressure control working distance (max sample height) 20mm -Approximately 350 degree free working area around crystal -Exchangable ATR crystal plate with automatic recognition of crystal typeand load of suitable mearurement parameters. -Easy exchange of sampling modules 3. Universal sampling module (A230/D) for transmittance measurements, -Spectral range of 375-7,500-1 -Sample compartment cover 2x3" standard sample mount 2x3" holder fro 13mm pellets

Bid #

Page 2 of 4

(Rev. 03/08)

REQUIREMENTS

INDICATE EXCEPTION OR STATE “AS SPECIFIED” BELOW

4. Gas Cell, 7cm (A129/D-11) -Body:ss316TI (1.4571), Window:ZnSe -Swagelok-connection to steel capillary 1/8" -o rings, Viton -Max pressure 25psi -2 replacement windows 5. Gas Cell 5cm (A132/D-11-H ZnSe windowa Pyrex body with 2 fittings for filling with replaceable rubber septa -O rings Viton -2 replaceable ZnSe windows (25x2mm), inclusive 2x3' holder

6. Opus Mentor Software OPUS Wizard: step-by-step assistant for the FT-IR analysis - Online help - Multimedia FT-IR tutorial - Measurement - Several functions for spectra displaying - Data pre-treatment, e.g. baseline correction, spectrum subtraction, atmospheric compensation - Peak Picking - Quick Compare - Spectrum interpretation - Starter library - Search spectra in IR-spectra library - Creation of user own libraries - Quantification (Lambert-Beer's Law) - Analyze peak areas and peak heights - Run, create and edit macros - Lab journal functionality - Predefined print layouts, customizable print layouts - Easy data export to other programs - Continuous monitoring of all spectrometer components - Multi level user management, indivi 7. Service Support -Warranty Extension to 10 Years for Interferometer Covers parts and labour in case of interferometer defects. The warranty starts with delivery of the spectrometer. Valid for ALPHA spectrometers. 8.Warranty Extension to 10 Years for laser Covers parts and labour in case of laser defects.

Bid #

Page 3 of 4

(Rev. 03/08)

REQUIREMENTS

INDICATE EXCEPTION OR STATE “AS SPECIFIED” BELOW

The warranty starts with delivery of the spectrometer. Valid for ALPHA spectrometers 9.Warranty Extension to 5 Years for MIR source Covers material costs in case of source defects. The warranty starts with delivery of the spectrometer. Valid for ALPHA spectrometers. 10.Warranty Extension to 10 Years for Platinum ATR A220/Dx or A225/Qx Covers material and labor costs in case of potentially necessary repairs in the Bruker factory. Defects resulting from improper operation are excluded from this. The warranty starts with the delivery of the accessory. 11. Consumable spar parts -MIR Source, pre-mounted, replaceable

pre-aligned,

user

12 .ATR-FTIR-Library COMPLETE (ATR-LIBCOMPLETE+) includes ATR spectra from polymers, monomers, additives, plasticizer, fillers, building materials, cosmetics, excipients, organic and inorganic chemicals, biochemicals, fibres, proteins, fatty acids, lipids, ingredients, natural products, silicon containing compounds, solvents, pesticides, pollutants, semiconductors, dyes, paints, coatings, food, food additives, minerals, lubricants, surfactants, kidney stones, pharmaceuticals and drugs. > 26,000 ATR-FTIR-Spectra; requires OPUS 7.2 or higher; only available with order of spectrometer 13.Valve for Gas cell 14. Transport panel for ALPHA spectrometer for battery pack C600-1/D (battery pack not included) Includes handle Compatible with slate PC

Bid #

Page 4 of 4

(Rev. 03/08)